gb
Cl-3100 Remote Sensor
FEATURES
Flow rate & Size Range
- Particles are sized greater than 0.5µm and 5µm. The standard flow rate is 1.0 cubic feet per minute, which limits the allowable concentration of particles to 1 million per cubic foot or 35.3 million per cubic meter. The sample volume can be collected in cubic foot mode (CF) or in cubic meter mode (CM). The sample time for the CF mode is 1 minute while the sample time for the CM mode is 35.3 minutes.
Power
- It is fully powered by a universal input power supply that will accept AC voltage from 100 to 240 VAC, 50/60 Hz. Two connections are made to the unit at its rear panel: a power cord (supplied with the unit), and a CAT 5 cable.
Design and Display
External Exhaust
- The CI-3100 has a filtered external exhaust from the unit. The exhaust port on the rear panel of the unit is connected to an internal filter that collects the sampled particles and any particles generated by the sampling system.
- There is an optional external exhaust tube fitting that can be installed so that the exhaust from the unit can be ported away from the unit or into an air return system. The tubing adapter is installed by unscrewing the exhaust hood from the rear panel and screwing in the tube fitting in its place.
Visual Indicators (Front)
- The CI-3100 provides five LED indicators on the front of the unit: Power, Alarm, Count, Laser Status and Flow Status. The POWER LED (green) illuminates when the power is turned on. The ALARM LED (red) illuminates if the flow or laser LED is off for 2 consecutive seconds, and will remain on until both the laser and flow measurements are back within their nominal range. The COUNT LED (yellow) flashes with every occurrence of a particle, up to ten counts per second.
- Higher concentrations will not flash any faster. The FLOW STATUS (green) LED is illuminated when there is flow through the sensor, measured with a mass flow sensor, and the flow rate is within ±10% of nominal. The LASER STATUS (green) LED is illuminated when the power level of the laser diode is within the factory set level ±10%. Both the flow and laser values are monitored 4 times per second, so their LEDs indicate in real time the flow and laser status.
Visual Indicators (Back)
- The CI-3100 provides two small LED indicators on the rear of the unit. These indicators are on the Ethernet communications module. A green LED will be on solid when the unit has just been turned on or when the Ethernet module is rebooting. The green LED will be flashing when there is network traffic in and out of the sensor. An amber/orange LED will be on when the Ethernet module senses a connection to an external network.
Output/communication
Ethernet MODBUS connects to data acquisition platform
- Connects easily to your current network infrastructure
- Built-in website with individual web address
- Choice of 3 web pages
- Operating data easily viewed by web browser
- Latch count or rolling count option
- Forced count feature simplifies validation
Others
- EU GMP Annex 1, ISO 144644-1, FS-209E calculation and printed Pass/Fail report
- Meets or exceeds ISO 21501-4
- Successfully brings ease of use and configurability to continuous monitoring
- Dramatically simplifies operations, including setup, installation, connection, communication, and validation
- Easy integration to your network system with many available communication options
- Readily connects to a variety of data acquisition or standard monitoring platforms
- Choice of internal built-in vacuum pump or external user-supplied vacuum source
- Built-in flow rate sensor available for fail-safe operation
- Laser status and flow measurement continuously monitored and controlled
- Designed to minimize particle emission (certified ISO class 2 level emission)
- Can be remotely controlled and monitored
- Easy to clean rugged seamless stainless steel enclosure and small footprint
Application
- Controlled Environments and Critical Processes
- Facility Certification for Compliance Regulations
- Continuous Monitoring of Clean rooms
- Indoor Air Quality
- Process Control
- Remote Sampling
- Self Diagnosis
- QA Programs
- Trend Tracking & Analysis
- Statistical Process Control
Standard Accessories
- 7-pin or 4-pin DIN connector (mating)
- User Manual
- AC Power Cord
- Quick Start Guide
- Setup CD-ROM
- 1/2” External Exhaust Fitting (if applicable)
cn
Cl-3100远程粒子计数器
特点
流速和粒子直径范围
- 测量的粒子直径是5µm和 5µm。标准流速是1立方英尺每分钟,测试浓度限度是每立方英尺100万个或每立方米3530万个。样品体积的单位可以是立方英尺或立方米。取1个立方英尺体积样品的时间是1分钟,取1立方米体积样品的时间是35.3分钟。
电源
- 通用输入电源:100伏特到240伏特,50/60 Hz交流电。后面板有两个接口:电源线(随设备提供)接口和CAT 5电缆接口
设计特点和显示器
外部排气
- CI-3100型远程粒子计数器 有排气过滤功能。位于后面板的排气口连接着一个内置过滤器,可以过滤取样气体的粒子和取样系统产生的粒子。
- 可选择使用外部排气管。排气管将设备的排气或回风系统。将后面板的排气罩拧下后,再安装排气管适配器。
可视指示灯(前面板)
- CI-3100远程粒子计数器的前部安装有5个LED指示灯:电源,报警,计数,激光管状态和流量状态。打开电源后,电源LED灯(绿色)变亮。 如果流量LED灯或激光管状态LED灯连续2秒熄灭,报警LED灯(红色)变亮,并且直到激光管状态和流量都回到正常状态后,报警LED灯才熄灭。 当有粒子通过计数器时,计数LED灯(黄色)开始闪烁。
- 更高的浓度不会导致更快的闪烁。当有气流通过质量流量传感器时,并且流速在正常值的±10%内,流量状态LED灯(绿色)变亮。当激光二极管的电压值在出厂设定值的±10%内,激光管状态LED灯(绿色)将变亮。流量和激光管的状态监测频率是每秒4次,保证LED灯实时指示流量和激光管状态。
可视指示灯(后面板)
- CI-3100远程粒子计数器的后面板配置了2个LED指示灯。这些指示灯安装在以太网接口模块上。当设备启动或当以太网模块重启时,绿色LED指示灯常亮。当传感器产生数据输入和输出,绿色LED灯将闪烁。当以太网模块连接到远程网络中,桔色LED灯变亮。
输出/通信
以太网MODBUS连接到数据采集平台
- 轻松连接到现有的网络
- 3个可选网页
- 通过网络浏览器查看运行数据
- 可选
- 强制计数功能,简化了验证工作
其它特点
- 满足EU GMP附录1,ISO 144644-1标准,FS-209E标准的计算方法和打印的报告
- 满足或超过ISO 21501-4标准
- 将持续监控的使用和配置变得容易
- 极大简化了使用,包括设置,安装,连接,通信和验证
- 可以选择多种通信接口,可以轻松连接到您的网络中。
- 轻松连接到各种数据采集或标准监控平台
- 可选择内置真空泵或使用用户提供的外部真空源
- 内置流量传感器,保证运行安全
- 持续监测和控制激光管状态和流量测量
- 使粒子产生水平降到最低的设计(满足ISO 2级水平)
- 可远程控制和监测
- 无缝不锈钢材质外壳方便清洁。设备仅产生微量粒子。
应用领域
- 受控环境和关键工艺
- 厂房的法规符合性认证
- 洁净室持续监控
- 室内空气质量检测
- 工艺控制
- 远程取样
- 自我诊断
- QA项目
- 趋势追踪和分析
- 统计工程控制
标准附件
- 7针或4针DIN接口
- 用户手册
- 交流电源线
- 快速使用指南
- 设置CD-ROM光盘
- 1/2”外部排气配件(可选)